A composition with three to the 797 connected Dosinos is recommended for the
Brightener determination with the Calibration technique MLAT:

Fig. 29:

Electrical connection for Brightener determination with the 838
Advanced Sample Processor and MLAT

A 50 mL Dosing Unit filled with VMS is connected to Dosino 1. A 2 mL Dosing Unit fil-
led with Brightener standard solution is connected to Dosino 2. A 2 mL Dosing Unit
filled with Suppressor concentrate is connected to Dosino 3.

Tubing connections

The choice of the tubing connections depends on the sample volume. Following 2 op-
tions are described:

Sample volume <10 mL in 11 mL sample vessels

Sample volume >10 mL in 50 mL sample vessels

Use thicker tubes for larger sample volumes.