Document logs – Bio-Rad Image Lab™ Software User Manual

Page 187

Advertising
background image

Document Logs

User Guide

| 187

6.

Enter a new name for the file and click Save.

Note:

Signed protocols are saved with an .sptl extension, signed images

are saved with an .sscn extension.

Document Logs

Any changes made or actions performed on a protocol or image file generate a

document log documenting each change or action. This log is created as soon as

the protocol or scan file is created. Protocols are updated when the protocol is

saved or run. If you make changes to a file and save it as a new file, regardless of

whether you are in secure mode, the document log is preserved in the new file. If

you are in secure mode, the signature of the previous file is noted as part of the

document log.

Note:

When a protocol is run, its log entries are copied to the resulting image

file.

All major actions and changes are audited (generate a document log). Examples of

auditable actions include:

Signing a file

Changing protocol settings

Changing the image, for example, by cropping or rotating it

Modifying the analysis, for example, changing the lanes and bands, or

adding or editing annotation to the images

Minor changes that affect only the display are not audited, such as:

Selecting different columns in the Analysis Table

Changing the display options in Lane Profile

Annotating or labeling the image using Annotation Tools

Each change you make to a signed protocol or image file must be documented in

the Reason for Change dialog box.