Application Tables

The following tables specify the light sources and conversion screens or filters you

need for each type of gel or blot to be imaged.

Detection
Reagent

Preferred
Light
Source

Alternate
Light
Source

Preferred
Filter

Alternate
Filter

Preferred
Flat Field

Alternate
Flat Field

Ethidium
bromide

UV

None

Standard

630/30

Orange

Lens

SYBR®
Green

XcitaBlue™

UV

560/50

Standard

Lens

None

SYBR® Safe

XcitaBlue

UV

560/50

Standard

Lens

None

SYBR® Gold

XcitaBlue

UV

560/50

Standard

Lens

None

GelGreen

XcitaBlue

UV

560/50

Standard

Orange

Lens

GelRed

UV

XcitaBlue

630/30

Standard

Lens

None

Fast Blast™

White trans

None

Standard

None

White
conv

Lens

Table 4.

Nucleic

Acids

Table 5.

Protein

Gels

Detection
Reagent

Preferred
Light
Source

Alternate
Light
Source

Preferred
Filter

Alternate
Filter

Preferred
Flat Field

Alternate
Flat Field

Coomassie
Blue

White trans

None

Standard

None

White
conv

Lens

Copper stain

White trans

None

Standard

None

White
conv

Lens

Zinc stain

White trans

None

Standard

None

White
conv

Lens

Flamingo™ UV

None

Standard

Standard

Orange

Lens

Oriole™

UV

None

Standard

None

Orange

Lens