A maintenance, Uv transilluminator lamp and starter replacement, Appendix a – Bio-Rad Image Lab™ Software User Manual

Page 189: Maintenance, Amaintenance

Advertising
background image

User Guide

| 189

A

Maintenance

This chapter includes instructions for maintaining the universal hood in proper

working condition by replacing parts.

UV Transilluminator Lamp and Starter Replacement

Note:

Keep the UV filter surface clean from the chemical agents used as gel

dyes. Use protective gloves when touching the UV transilluminator cover.

Depending on usage, the UV bulbs and starters last for many years. Replace bulbs

when you notice them flickering. If a bulb does not turn on when it is new or moved,

replace the bulb starter and test the bulb again.

There are three types of bulbs available. The catalog numbers are listed in

Appendix

D, Accessories

. The standard bulb is 302 nm. Optionally, the 254 nm bulb is used

for cross-linking of proteins and nucleic acids, and the 365 nm bulb is used to

minimize denaturing of DNA.

To replace the lamps

1.

Turn off the power.

2.

Disconnect the power cord from the universal hood.

3.

Remove the four screws located on the left/right sides of the transilluminator
cover.

4.

Remove the cover with the UV glass by sliding it forward, then lifting up.

Advertising