A

| Maintenance

190

| Gel Doc XR+ and ChemiDoc XRS+ Systems with Image Lab Software

5.

Place it on a nonabrasive surface so that the glass does not get scratched or
damaged.

Note:

Do not put the UV cover directly on the bench. Wear gloves when

touching the lamps.

6.

Rotate the lamp until it becomes loose and the pins come to a vertical position.

7.

Remove the lamp. Install the new lamp by rotating it so that the pins are
horizontal and the lamp is tight.

8.

Remove the starter by rotating it counterclockwise and then pulling it out.

9.

Insert the new starter into the holder and rotate clockwise.

10.

.

Reassemble the cover and retighten the screws on both sides.