Epi-Illumination Lamp Replacement

User Guide

| 191

Epi-Illumination Lamp Replacement

The lamps are located behind two cover panels on the left and right sides of the

universal hood’s interior.

To replace the lamps

1.

Turn system power off and remove the power cord.

2.

Open the enclosure door.

3.

Locate the epi-white light housing inside the hood and the screw on the center
of the housing cover.

4.

Use a socket wrench to remove the screw.

5.

Pull the cover to remove it from the housing and the lamp will become visible.

6.

To remove the lamp, hold it at the plastic receptacle and pull it from the
receptacle.

7.

Insert the new lamp into the lamp receptacle and push until it is firmly seated in
the receptacle.

8.

Reassemble the epi-white light cover.