3 front panel configuration, Front panel configuration, Section 3.1.3 on – Rice Lake 420 Plus HMI Digital Weight Indicator Installation Manual User Manual

Page 20

Advertising
background image

16

420 Plus Installation Manual

3.1.3

Front Panel Configuration

The

420 Plus

indicator can be configured using a series of menus accessed through the indicator front panel when

the indicator is in setup mode. Table 3-1 summarizes the functions of each of the main menus.

Table 3-1. 420 Plus Menu Summary

Menu

Menu Function

CONFIG

Configuration

Configure grads, zero tracking, zero range, motion band, overload, tare function, sample rate,
and digital filtering parameters.

FORMAT

Format

Set format of primary and secondary units, display rate.

CALIBR

Calibration

Calibrate indicator. See Section 4.0 on page 30 for calibration procedures.

SERIAL

Serial

Configure EDP and printer serial ports.

PROGRM

Program

Set power-up mode, regulatory mode, and consecutive number values.

PFORMT

Print Format

Set print format used for gross and net tickets. See Section 6.0 on page 39 for more
information.

SETPNT

Setpoint

Configure Setpoints and digital outputs.

DIG IN

Digital Input

Assign digital input functions.

ALGOUT

Analog Output

Configure analog output module. Used only if analog output option is installed.

VERS

Version

Display installed software version number.

Move UP/

Increrement

VAlue

ENTER value

Move LEFT/

Previous

Move RIGHT/

Next

Move DOWN/

Decrement

Value

Figure 3-2. Front Panel Key Functions in Setup Mode

Four front panel keys are used as directional keys to navigate through the menus in setup mode (see Figure 3-2).

The

UNITS

( ) and

PRINT

( ) keys scroll left and right (horizontally) on the same menu level;

ZERO

( ) and

GROSS/NET

( ) move up and down (vertically) to different menu levels. The

TARE

key ( ) serves as an Enter key

for selecting parameter values within the menus. A label above each of these keys identifies the direction provided

by the key when navigating through the setup menus.

1

st

Level

Parameter

Default value

Value

When moving thr ough values below the first menu level, press

to r etur n to the level

above. Press

or

to move to the next parameter on the level above

1

st

Level

Parameter

2

nd

Level

Parameter

2

nd

Level

Parameter

Value

Value

Figure 3-3. Setup Mode Menu Navigation

Advertising