2 reporting commands, 3 the resetconfiguration command, 4 parameter setting commands – Rice Lake 420 Plus HMI Digital Weight Indicator Installation Manual User Manual

Page 38: 5 soft reset, Reporting commands, The resetconfiguration command, Parameter setting commands, Soft reset

Advertising
background image

34

420 Plus Installation Manual

5.1.2

Reporting Commands

Reporting commands (see Table 5-2) send specific

information to the EDP port. These commands can be

used in both setup mode and normal mode.

Table 5-2. EDP Reporting Commands

Command

Function

DUMPALL

List all parameter values

VERSION

Write

420 Plus

software version

P

Write current displayed weight with units
identifier. See Section 7.2 on page 43 for
more information.

ZZ

Write current weight and annunciator
status. See Section 7.2 on page 43 for
more information.

S

Write one frame of stream format. See
Table 5-12 on page 37 for information
about the SX and EX commands.

5.1.3

The RESETCONFIGURATION Command

The RESETCONFIGURATION command can be used

to restore all configuration parameters to their default

values. Before issuing this command, the indicator

must be placed in test mode (press and hold setup

switch for approximately three seconds to show TEST

menu).
This command is equivalent to using the DEFLT

function on the TEST menu. See Section 7.9 on

page 51 for more information about test mode and

using the TEST menu.

Note

All load cell calibration settings are lost when
the RESETCONFIGURATION command is run.

5.1.4

Parameter Setting Commands

Parameter setting commands allow you to display or

change the current value for a particular configuration

parameter (Tables 5-3 through 5-11).
Current configuration parameter settings can be

displayed in either setup mode or normal mode using

the following syntax:

command<ENTER>

Most parameter values can be changed in setup mode

only. Use the following command syntax when

changing parameter values:

command=value<ENTER>

where

value

is a number or a parameter value. Use no

spaces before or after the equal (=) sign. If you type an

incorrect command or value, the display reads

??

.

Changes to the parameters are saved as they are entered

but typically do not take effect until you exit setup

mode.
For example, to set the motion band parameter to 5,

type the following:

MOTBAND=5D<ENTER>

To view choices, type the following:

MOTBAND=?<ENTER>

5.1.5

Soft Reset

The command RS will reset the indicator without

losing configuration and calibration. It does reset

consecutive number and count mode.

Table 5-3. CONFIG EDP Commands

Command

Description

Values

GRADS

Graduations

1–100 000

ZTRKBND

Zero track band

OFF, 0.5D, 1D, 3D

ZRANGE

Zero range

1.9%, 100%

MOTBAND

Motion band

1D, 2D, 3D, 5D, 10D, 20D, OFF

OVRLOAD

Overload

FS+2%, FS+1D, FS+9D, FS

SMPRAT

Sample rate

15HZ, 30HZ, 60HZ, 7.5HZ

DIGFLTR1
DIGFLTR2
DIGFLTR3

Digital filtering

1, 2, 4, 8, 16, 32, 64

DFSENS

Digital filter cutout sensitivity

2OUT, 4OUT, 8OUT, 16OUT, 32OUT, 64OUT, 128OUT

DFTHRH

Digital filter cutout threshold

NONE, 2DD, 5DD, 10DD, 20DD, 50DD, 100DD, 200DD, 250DD

TAREFN

Tare function

BOTH, NOTARE, PBTARE, KEYED