5 program menu, Program menu, Figure 3-9. program menu – Rice Lake 420 Plus HMI Digital Weight Indicator Installation Manual User Manual

Page 27: Table 3-5. serial menu parameters (continued), Table 3-6. program menu parameters

Advertising
background image

420 Plus Installation Manual - Configuration

23

3.2.5

Program Menu

XXXXXXX

XXXXXXX

PWRUPM

GO

OIML

REGULA

000000

number

CONSNU

DELAY

SETPNT

PROGRM PFORMT

SERIAL

CALIBR

CONFIG FORMAT

000000

number

CONSTU

NTEP

CANADA

1

UID

VERS

DIGIN ALGOUT

DATFMT

MMDDYY

DATE

DDMMYY

YYMMDD

YYDDMM

DATSEP

SLASH

DASH

SEMI

TIMFMT

12HOUR

TIME

24HOUR

TIMSEP

COLON

COMA

NONE

COUNT

ACCESS

SPLISIZ

ACCUM

ON/OFF

ON

ON

RTZGRD

0

Figure 3-9. Program Menu

S BITS

1 STOP
2 STOP

Stop bits. Sets the number of stop bits to 1 or 2.

TERMIN

CR/LF
CR

Termination character. Selects termination character for data sent from the EDP or printer port.

EOLDLY

000000
number

End-of-line delay. Sets the delay period, in 0.1-second intervals, from when a formatted line is
terminated to the beginning of the next formatted serial output. Value specified must be in the
range 0-255, in tenths of a second (10 = 1 second).

Note

An EOL may be required for continuous transmission at slower baud rates to
ensure the receiving buffer is empty before another string is transmitted

ECHO

OFF
ON

This command enables or disables echoing of the serial commands sent to the indicator.

SERIAL Menu

Parameter

Choices

Description

Table 3-5. Serial Menu Parameters (Continued)

PROGRM Menu

Parameter

Choices

Description

Level 2 submenus

PWRUPM

GO
DELAY

Power up mode. In GO mode, the indicator goes into operation immediately after a brief power
up display test.

In DELAY mode, the indicator performs a power up display test, then enters a 30-second
warm up period. If no motion is detected during the warm-up period, the indicator becomes
operational when the warm up period ends; if motion is detected, the delay timer is reset and
the warm up period repeated.

COUNT

ACCESS
SPLSIZ

Specifies whether operator has access to piece count mode and the default sample size used
for parts counting. See level three submenu for parameter descriptions.

Table 3-6. Program Menu Parameters

Advertising