420 Plus Installation Manual

4.0

Calibration

The

420 Plus

can be calibrated using the front panel, EDP commands, or the

Revolution

®

configuration utility.

Each method consists of the following steps:

Zero calibration

Entering the test weight value

Span calibration

Optional rezero calibration for test weights

using hooks or chains.

The following sections describe the calibration procedure for each of the calibration methods.

XXXXXXX

XXXXXXX

XXXXXXX

XXXXXXX

WZERO

WV AL

WSPAN

WLIN

ALGOUT

DIGI N

SETPNT

PROGRM PFORMT

SERIAL

CALIBR

CONFIG FORMAT

REZERO

VERS

WPT–>

1

WPT–>

2

WPT–>

3

WPT–>

4

WPT–>

5

Same as PT ->

1

*

CAL*

Display and edit
span calibration
A/D count value

Figure 4-1. Calibration (CALIBR) Menu

4.1

Front Panel Calibration

To calibrate the indicator using the front panel,

1. Place the indicator in setup mode (display

reads

CONFIG

) and remove all weight from the

scale platform. If your test weights require

hooks or chains, place the hooks or chains on

the scale for zero calibration.

2. Press until the display reads

CALIBR

(see

Figure 4-1). Press to go to zero calibration

(

WZERO

).

3. With

WZERO

displayed, press to calibrate

zero. The indicator displays

*CAL*

while

calibration is in progress. When complete, the

A/D count for the zero calibration is displayed.

DO NOT adjust this value after WSPAN has

been set! Press again to save the zero

calibration value and go to the next prompt

(

WVAL

).

4. With

WVAL

displayed, press to show the

stored calibration weight. Use the procedure

shown in Figure 3-4 15 to enter the actual

value of the test weights to be used for the span

calibration. Press to store the entered

WVAL

value and go to span calibration (

WSPAN

).

5. With

WSPAN

displayed, place test weights on

the scale and press to calibrate span. The

indicator displays

*CAL*

while calibration is in

progress. When complete, the A/D count for

the span calibration is displayed. Press again

to store the span calibration value and go to the

next prompt (

REZERO

).

6. Five-point linearization (using the WLIN

parameter) provides increased scale accuracy

by calibrating the indicator at up to five

additional points between the zero and span

calibrations.
Linearization is optional: if you choose not to

perform linearization, skip the WLIN

parameter; if linearization values have

previously been entered, these values are reset

to zero during calibration. To perform

linearization, follow the procedure below:
Standard Calibration: With

WLIN

displayed,

Press to go to the first linearization point

(

PT-> 1

). Place test weights on the scale and

press or

ENTER

.

Enter the actual test weight value, then press

to calibrate. The indicator displays

*CAL*

while

calibration is in progress. When complete, the

A/D count for the linear calibration is

displayed. Press again to save the calibration

value and go to the next prompt (

PT-> 2

).

Repeat for up to five linearization points. To

exit the linearization parameters, press to

return to WLIN.
Millivolt Calibration: If millivolt calibration is

enabled press to display and edit the test

weight value, press again to display and edit

the millivolt value for that weight, press

again to calibrate and display the raw A/D

count, press again to move to the next

calibration point.