420 Plus Installation Manual - Configuration

21

3.2.3

Calibration Menu

See Section 4.0 on page 30 for Calibration procedures.

XXXXXXX

XXXXXXX

XXXXXXX

XXXXXXX

WZERO

WV AL

WSPAN

WLIN

ALGOUT

DIGI N

SETPNT

PROGRM PFORMT

SERIAL

CALIBR

CONFIG FORMAT

REZERO

VERS

WPT–>

1

WPT–>

2

WPT–>

3

WPT–>

4

WPT–>

5

Same as PT ->

1

*

CAL*

Display and edit
span calibration
A/D count value

Figure 3-7. Calibration Menu

Table 3-4. Calibration Menu Parameters

CALIBR Menu

Parameter

Choices

Description

Level 2 submenus

WZERO

Display and edit the zero calibration A/D count value.

DO NOT adjust this value after WSPAN has been set!

WVAL

Display and edit the test weight value.

WSPAN

Display and edit the span calibration A/D count value.

WLIN

WPT->1 —
WPT->5

Press ENTER to display and edit test weight value. Pressing ENTER again will calibrate and
display the raw A/D value. Pressing ENTER a third time will move to the next calibration point.

For millivolt calibration, press ENTER to display and edit the test weight value. Press ENTER
again to display and edit the millivolt value for that weight. Press ENTER a third time to
calibrate and display the raw A/D value. Press ENTER the fourth time to move to the next
point.

REZERO

Press Enter to remove an offset value from the zero and span calibrations if hooks or chains
are being used during calibration.

Always use this parameter after WZERO and WSPAN have been set to re-capture a new zero
value. See Section 4.1 on page 30 for more information about using this parameter.