4 more about calibration, 1 adjusting final calibration, 2 zero deadload a/d counts – Rice Lake 420 Plus HMI Digital Weight Indicator Installation Manual User Manual

Page 36: Adjusting final calibration, Zero deadload a/d counts

Advertising
background image

32

420 Plus Installation Manual

6. When span calibration is complete, a dialog box

asks whether you wish to perform linear

calibration. Click

Ye s

to perform linear

calibration for up to five linearization points, or

click

No

to complete calibration (see step 8).

7. On the Linear Calibration display, select the

point (1–5) to calibrate, then click

Calibrate

.

Place test weights on scale then click

OK

.

When prompted, enter the test weight value

then press Enter. Repeat for additional

linearization points, then click

Exit

to return to

the Indicator Calibration display.

8. When calibration is complete, the

New Settings

fields of the Indicator Calibration display are

filled in. Click

Exit

to save the new values and

return to the

Revolution

main menu; to restore

the previous calibration values, click

Restore

Settings

.

4.4

More About Calibration

The following topics provide additional information

about compensating for environmental factors

(Section 4.4.1) and diagnostic information for

determining expected zero and span coefficients.

4.4.1

Adjusting Final Calibration

Calibration may be affected by environmental factors

including wind, vibration, and angular loading. For

example, if the scale is calibrated with 1000 lb, a strain

test may determine that at 2000 lb the calibration is 3 lb

high. In this case, final calibration can be adjusted by

tweaking WVAL to 998.5 lb. This adjustment provides

a linear correction of 1.5 lb per 1000 lb.
To adjust the final calibration, return to the

WVAL

prompt and press to show the test weight value.

Press or to adjust calibration up or down. Press

to save the value, then press to return to the CALIBR

menu.

4.4.2

Zero Deadload A/D Counts

Table 4-1 lists the ideal A/D counts that result from

input signals of 0–3.0 mV with zero deadload. Actual

values will typically be higher than the values shown in

Table 4-1 but the ideal values can be used when

calibrating the indicator with no attached scale.

Table 4-1. Ideal A/D Raw Counts

Input Signal (mV)

Raw A/D Counts

0.0 mV/V

167,840

0.5 mV/V

335,613

1.0 mV/V

503,377

1.5 mV/V

671,143

2.0 mV/V

838,908

2.5 mV/V

1,100,668

3.0 mV/V

1,174,446

Note

When Raw A/D Count is displayed, the six most
significant digits appear. Scroll left or right to
see the full value.