420 Plus Installation Manual - Calibration

31

7. The rezero function is used to remove a

calibration offset when hooks or chains are

used to hang the test weights.

• If no other apparatus was used to hang the test

weights during calibration, remove the test

weights and press to return to the CALIBR

menu.

• If hooks or chains were used during

calibration, remove these and the test weights

from the scale. With all weight removed, press

to rezero the scale. This function adjusts the

zero and span calibration values. The indicator

displays

* C A L *

while the zero and span

calibrations are adjusted. When complete, the

adjusted A/D count for the zero calibration is

displayed. Press to save the value, then press

to return to the CALIBR menu.

8. Press until the display reads

CONFIG

, then

press to exit setup mode.

4.2

EDP Command Calibration

To calibrate the indicator using EDP commands, the

indicator EDP port must be connected to a terminal or

personal computer. See Section 2.3.5 on page 10 for

EDP port pin assignments; see Section 5.0 on page 33

for more information about using EDP commands.
Once the indicator is connected to the sending device,

1. Place the indicator in setup mode (display

reads

CONFIG

) and remove all weight from the

scale platform. If your test weights require

hooks or chains, place them on the scale for

zero calibration.

2. Send the WZERO EDP command to calibrate

zero. The indicator displays

*CAL*

while

calibration is in progress.

3. Place test weights on the scale and use the

WVAL command to enter the test weight value

in the following format:

WVAL=nnnnnn<CR>

4. Send the WSPAN EDP command to calibrate

span. The indicator displays

*CAL*

while

calibration is in progress.

5. Up to five linearization points can be

c a l i b r a t e d b e t w e e n t h e z e r o a n d s p a n

c a l i b r a t i o n v a l u e s . U s e t h e f o l lo w i ng

commands to set and calibrate a single

linearization point:

WLIN.V1=nnnnn<CR>
WLIN.C1<CR>

The WLIN.V1 command sets the test weight

value (nnnnn) for linearization point 1. The

WLIN.C1 command calibrates the point.

Repeat using the WLIN.Vx and WLIN.Cx

commands as required for additional

linearization points.

6. To remove an offset value, clear all weight

from the scale, including items used to hang

test weights, then send the REZERO EDP

command. The indicator displays

*CAL*

while

the zero and span calibrations are adjusted.

7. Send the KUPARROW EDP command to exit

setup mode.

4.3

Revolution Calibration

To calibrate the indicator using

Revolution

, the

indicator EDP port must be connected to a PC running

the

Revolution

configuration utility.

Use the following procedure to calibrate the indicator:

1. Select

Calibration Wizard

from the

Revolution

tools menu.

2.

Revolution

uploads calibration data from the

indicator then presents the information in the

calibration wizard, as shown in Figure 4-2

Figure 4-2. Revolution Calibration Display

3. Enter the

Value of Test Weight

to be used for span

calibration then click

OK

.

4. The Zero Calibration dialog box prompts you

to remove all weight from the scale. Clear the

scale and click

OK

to begin zero calibration.

Note

If your test weights require hooks or
chains, place them on the scale for zero
calibration.

5. When zero calibration is complete, the Span

Calibration dialog box prompts you to place

test weights on the scale for span calibration.

Place tests weights on the scale then click

OK

.