Rice Lake Counterpart Configurable Counting Indicator - User Manual - Version 1.0 User Manual

Page 31

Advertising
background image

Counterpart User Manual - Configuration

25

FEATUR Menu

Parameter

Choices

Description

Level 3 submenus - Feature

COUNT

NEGCOUNT
SAMPLEQTY
INSFSMPL
LOTUPDT
SCLCHG
NEWITEM
DISPACC

UNITWTUPDATE

UTWTBASE

Selects the counting mode (enables/disables negative checkweighing; turns data parameters
on/off. See Level 3 submenus.

REGION

REGULA
REGWRD
DECFMT
TIME
DATE

Selects regional settings. See Level 3 submenus.

UID

1

Sets the unit ID, a string of up to 6 ASCII characters, which can be set via serial port or
keypad. This will be used in place of the <UID> token in a print format. The default value is “1.”

CONSC#

CURVAL
RESVAL

Allows sequential numbering for print operations (CURVAL is the current value and RESVAL is
the reset value). The consecutive number value is incremented following each print operation
that includes <CN> in the ticket format. When the consecutive number is reset, it is reset to
the RESVAL specified on the parameter.

PASSWD

CALIBR
SETUP
TEST
T&D
ID

Creates a password to access the CALIBR, SETUP, TEST, T&D, and ID menus. Specify a
non-zero value to restrict access to all configuration menus. Passwords can be overridden by
loading new firmware or entering 999999.

Overriding passwords will clear configuration and calibration settings. To
preserve settings (i.e., ID information), use Revolution software to upload
your data to a PC, then download it back to the indicator after the
password override is performed.

SOFTKEYS

SK1-12

KEYLCK

MENU
ZERO
GRSNET
UNITS
PRINT
TARE
NUMBER
SAMPLE
UNIT WT

Disables the MENU, ZERO, GRSNET, UNITS, PRINT, TARE number sample and unit weights.
Select LOCK to disable the key, and UNLOCK to enable the key.

STNDBY

0

Standby mode delay. Specifies the number of minutes the indicator must be inactive before
entering standby mode. Valid values are 0 (off) through 255 minutes. When in standby mode,
power is still supplied to the CPU and draws 1/2 of the current as when the display is
powered. The annunciators are still lit, but no weight is displayed. Press any key to exit
standby mode and reactivate the display.

The indicator enters standby mode if no key presses, serial communications, or scale motion
occur for the length of time specified in this parameter. Set to 0 to disable standby mode.

Standby mode will not be implemented if there is weight on the scale.

Table 3-7. FEATUR Submenu Parameters

Note

Note