Counterpart User Manual - Calibration

39

4.0

Calibration

The

Counterpart

can be calibrated using the front panel, EDP commands, or Revolution

®

.

Calibration can be performed in two places within the menu: the CALIBR menu shown in Figure 4-1 and the
SCALE submenu shown in Figure 3-6 on Page 18. The CALIBR menu shown in Figure 4-1 is a “quick
access” calibration; for more in-depth scale setup and calibration, use the menus found under
SETUP

»

CONFIG

»

SCALE (see Figure 3-6 on Page 18).

Calibration consists of the following steps:

Zero calibration

Entering the test weight value

Span calibration

Optional five-point linearization

Optional rezero calibration for test weights
using hooks or chains.

The

Counterpart

requires the WZERO and WSPAN points to be calibrated. The linearity points are

optional, but must NOT duplicate zero or span. During calibration, the TARE(Enter) key acts as a data entry
confirmation key. The soft calibrate key accepts that value.

Figure 4-1. Calibration (CALIBR) Menu

4.1

Front Panel Calibration

1. Press the

MENU SETUP

button, then press

u n t i l t h e d i s p l a y r e a d s

C A L I B R

( s e e

Figure 4-1).

2. Press

TARE(Enter)

or to go to the scale to be

calibrated. Press

TARE(Enter)

or to go to

WZERO

. Press

TARE(Enter)

once again to

display a value and press the

Calibrate softkey

to

Calibrate

WZERO.

If you do not want to calibrate zero, press
Menu to exit.

3. The indicator displays

CALIBRATING

while

calibration is in progress. When complete,

press

TARE(Enter)

or to advanced to the next

prompt (

WVAL

).

To view the new A/D count, repeat Step 3,
but instead of pressing TARE(Enter) while
viewing the value, press MENU SETUP to

exit.

4. With

WVAL

displayed, press

TARE(Enter) or

to display or edit the stored calibration weight
value. Press

TARE(Enter)

to store the

WVAL

value and advance to

WSPAN

.

5. With

WSPAN

displayed, press

TARE(Enter)

or

to view the A/D count. Place test weights on
the scale equal to WVAL. Press the

Calibrate

softkey

to calibrate WSPAN.

If you do not want to calibrate span, press
Menu to exit.

6. After

Calibrate softkey

is pressed, the indicator

displays

CALIBRATING

when complete. Press

TARE(Enter)

or and the

SaveExit softkey

or

proceed to

WLIN

.

To view the new A/D count, navigate back to
the WSPAN, repeat Step 6; however, instead
of pressing Calibrate softkey while viewing

the value, press Save/Exit to exit.

Note

Note

WZERO

Calibrating

Calibrating

Display and edi

Press the calibrate

softkey to calibrate

zero.

t

test weight value

WVAL

WSPAN

Press calibrate

softkey to remove

offset from zero from

zero and span

calibrations

REZERO

WLI N

POINT 1

POINT 2

POINT 3

POINT 4

POINT 5

Same as POINT 1

WGT 1

CAL1

Calibrating

Press the calibrate

softkey to calibrate

span.

XXXXXXX

ID

DISPLAY

TEST

SETUP

AUDIT

CALIBR

Press Menu

to cancel

Press Menu to

cancel

Previous A/D raw

counts are shown

Previous A/D raw

counts are shown

SCALE 1

SCALE 2

SCALE 3

COMING SOON

Note

Note

Note