Rice Lake Counterpart Configurable Counting Indicator - User Manual - Version 1.0 User Manual
Page 75
Counterpart User Manual - For More Information
69
<M[G | N | T]>
STR.GROSS
STR.NET
STR.TARE
Mode. Specifies gross, net, or tare weight for the current or
specified weight on the source scale.
<S>
STR.MOTION
STR.RANGE
STR.OK
STR.INVALID
Status for the source scale. Default values and meanings for each
status:
STR.MOTIONMIn motion
STR.RANGEOOut of range
STR.OK<space>OK
STR.INVALIDIInvalid
<CW>
—
Status of over/under/accept. O=Over, U=Under, A=Accept.
<B [–]n,…>
See descriptions below
Bit fields. Comma-separated sequence of bit field specifiers. Must
be exactly 8 bits. Minus sign ([–]) inverts the bit.
B0
—
Always 0
B1
—
Always 1
B2
Configuration
=1 if even parity
B3
Dynamic
=1 if MODE=NET
B4
Dynamic
=1 if COZ
B5
Dynamic
=1 if standstill
B6
Dynamic
=1 if gross negative
B7
Dynamic
=1 if out of range
B8
Dynamic
=1 if secondary/tertiary
B9
Dynamic
=1 if tare in system
B10
Dynamic
=1 if tare is keyed
B11
Dynamic
=00 if MODE=GROSS
=01 if MODE=NET
=10 if MODE=TARE
=11 (not used)
B12
Dynamic
=00 if UNITS=PRIMARY
=01 if UNITS=SECONDARY
=10 if UNITS=TERTIARY
=11 (not used)
B13
Configuration
=00 (not used)
=01 if current DSPDIV=1
=10 if current DSPDIV=2
=11 if current DSPDIV=5
B14
Configuration
=00 (not used)
=01 if primary DSPDIV=1
=10 if primary DSPDIV=2
=11 if primary DSPDIV=5
B15
Configuration
=00 (not used)
=01 if secondary DSPDIV=1
=10 if secondary DSPDIV=2
=11 if secondary DSPDIV=5
B16
Configuration
=00 (not used)
=01 if tertiary DSPDIV=1
=10 if tertiary DSPDIV=2
=11 if tertiary DSPDIV=5
Format Identifier
Defined By
Description
Table 9-3. Custom Stream Formatting