Counterpart User Manual - Calibration

41

4.4

More About Calibration

The following topics provide additional information
about compensating for environmental factors
(Section 4.4.1) and diagnostic information for
determining expected zero and span coefficients.

4.4.1

Adjusting Final Calibration

Calibration may be affected by environmental factors
including wind, vibration, and angular loading. For
example, if the scale is calibrated with 1000 lb, a
strain tes t may deter m i ne t hat at 2 0 00 lb th e
calibration is 3 lb high. In this case, final calibration
can be adjusted by tweaking WVAL to 998.5 lb. This
adjustment provides a linear correction of 1.5 lb per
1000 lb.

To adjust the final calibration, return to the

WVAL

prompt and press

TARE(Enter)

to show the test weight

value. Press or to adjust calibration up or down.
Press

Calibrate softkey

to save the value, then press to

return to the CALIBR menu.