Counterpart User Manual - Calibration
41
4.4
More About Calibration
The following topics provide additional information
about compensating for environmental factors
(Section 4.4.1) and diagnostic information for
determining expected zero and span coefficients.
4.4.1
Adjusting Final Calibration
Calibration may be affected by environmental factors
including wind, vibration, and angular loading. For
example, if the scale is calibrated with 1000 lb, a
strain tes t may deter m i ne t hat at 2 0 00 lb th e
calibration is 3 lb high. In this case, final calibration
can be adjusted by tweaking WVAL to 998.5 lb. This
adjustment provides a linear correction of 1.5 lb per
1000 lb.
To adjust the final calibration, return to the
WVAL
prompt and press
TARE(Enter)
to show the test weight
value. Press or to adjust calibration up or down.
Press
Calibrate softkey
to save the value, then press to
return to the CALIBR menu.