EDP Command Calibration

To calibrate the indicator using EDP commands, the
indicator EDP port must be connected to a terminal or
personal computer. See Section 2.7.1 on page 8 for
EDP port pin assignments.
Once the indicator is connected to the sending device,
do the following:

1. Place the indicator in config mode (display

must read

SCALE

- see

Figure 3-6 on page 19)

a n d r e m o v e a l l w e i g h t f r o m t h e s c a l e
platform. If your test weights require hooks or
chains, place the hooks or chains on the scale
for zero calibration.

2. Send the SC.WZERO#n command to

c a l i b r a t e z e r o . T h e i n d i c a t o r d i s p l a y s

CALIBRATING

while calibration is in progress.

3. Place test weights on the scale and use the

SC.WVAL#n command to enter the test
weight value in the following format:

SC.WVAL=nnnnnn<CR>

4. Send the SC.WSPAN#n command to calibrate

span. The indicator displays

CALIBRATING

while calibration is in progress.

5. Up to five linearization points can be

c a l i b r a t e d b e t w e e n t h e z e r o a n d s p a n
c a l i b r a t i o n v a l u e s . U s e t h e f o l l o w i n g
commands to set and calibrate a single
linearization point:

SC.WLIN.V1=nnnnn<CR>
SC.WLIN.C1<CR>

The SC.WLIN.V1#n command sets the test
weight value (nnnnn) for linearization point 1.
The SC.WLIN.C1#n command calibrates the
point. Repeat using the SC.WLIN.Vx and
SC.WLIN.Cx commands as required for
additional linearization points.

6. To remove an offset value, clear all weight

from the scale, including hooks or chains used
t o h a n g t e s t w e i g h t s , t h e n s e n d t h e
SC.REZERO#n command. The indicator
displays

CALIBRATING

while the zero and span

calibrations are adjusted.

7. Send the SC.KUPARROW#n or

SC.KEXIT#n EDP command to return to
normal mode.

4.3

Revolution

®

Calibration

To calibrate the indicator using Revolution, the
indicator EDP port must be connected to a PC running
the Revolution configuration utility.

1. Place the indicator in config mode (display

reads

SCALE

, see

Figure 3-6 on page 19) and

remove all weight from the scale platform.

2. From Revolution, select

File

»

New

.

The Select Indicator dialog box appears.

3. Select

Counterpart

and click

OK

.

4. From the Communications menu, select

Connect

.

5. From the left pane, expand the Scale selection

and click the

Scale

button.

Figure 4-2. Scale Button

6. From the Tools menu, select

Calibration Wizard

.

7. Click

NEXT

to begin the Calibration Wizard.

8. Select whether you want to perform a

standard calibration or a standard with
multi-point linearization and click

NEXT

.

9. In the text box, enter the test weight value be

used for span calibration.

10. Select the checkbox if you are using chains or

hooks during the calibration, then click

NEXT.

11. Remove all weight from the scale and click

Click to Calibrate Zero

to begin zero calibration.

If your test weights require hooks or chains,
place them on the scale for zero calibration.

12. When zero calibration is complete, the

Calibration Wizard prompts you to place test
weights on the scale. Place the test weights on
the scale, then click

Click to Calibrate Span

.

13. If you chose to perform linear calibration, the

Calibration Wizard now displays prompts
(1–5). Enter the weight value for Linear Point
#1, place test weights on scale and click

GO

.

Repeat for additional linearization points,
then click

NEXT

.

14. The new and old calibration settings are

displayed. To accept the new values, click

Finish

. To exit and restore the old values, click

Cancel

.