3 continuous output (stream) format, 4 demand output serial data format, 5 custom stream formatting – Rice Lake Counterpart Configurable Counting Indicator - User Manual - Version 1.0 User Manual

Page 74

Advertising
background image

68

Counterpart User Manual

9.3

Continuous Output (Stream) Format

Figure 9-1 shows the continuous output format sent to a

Counterpart

port when that port’s TRIGGE parameter

(SERIAL menu) is set to STREAM.

Figure 9-1. Continuous Output Data Format

9.4

Demand Output Serial Data Format

When demand mode is configured for the serial port (STREAM parameter set to OFF), the

Counterpart

uses a

data string formatted for a basic ticket printout. The particular ticket format printed depends on the indicator
configuration.

You can customize the ticket to work with a wide variety of printers, scoreboard displays, and other remote
equipment. See Section 6.0 on page 52 for more information on custom print formats.

9.5

Custom Stream Formatting

Each port can be independently configured to stream a default frame format or can be customized to stream a
user-defined format. Custom formatting is very similar to the standard print formatting described in Section 7.0
on page 52
. Custom stream formatting can be set up through the Regula/Indust submenu. Table 9-3 lists the
format identifiers used to configure a custom stream format.

<STX> <POL> <wwwwwww> <UNIT> <G/N> <S> <TERM>

ASCII 02

(decimal)

Weight data: 7 digits, right-justified, with

decimal point, l eading zer o s upr e ssion.

Overload = &&&&&&&

Underrange = = = = = = = =

Display overflow = OVERFL

L = pounds

K = kilogram
G = grams

O = ounces

G = Gross

N = Net

Status:

< space > = valid

I = Invalid

M = Motion

O = Over/under range

Polarity:

< space > = Positive

<–> = Negative

<^> = Overload

<]> = Underrange

<CR> <LF>

or <CR>

ASCII 13, 10

(decimal)

<space> = lb/oz

Format Identifier

Defined By

Description

<P[G | N | T]>

STR.POS
STR.NEG

Polarity. Specifies positive or negative polarity for the current or
specified (Gross/Net/Tare) weight on the source scale. Possible
values are SPACE, NONE, + (for STR.POS), or – (for STR.NEG)

<CU>

STR.PRI
STR.SEC
STR.TER

Units. Specifies primary, secondary, or tertiary units for the current or
specified weight on the source scale.

<U[P | S | T]>

Dynamic

=L if Units=LB
=K if Units=KG
=G if Units=G
=O if Units=OZ
=space if Units=LB/OZ

Table 9-3. Custom Stream Formatting

Advertising