Trade

Route Installation Manual

Figure 2-23. Junction Box Indicator Terminal

2.13 920i Indicator Junction Box Cable Installation

The homerun cable has been installed by the factory. If required, attach cable from a junction box to an installed A/

D card, route the cable through the cord grip and ground the shield wire.
Next, remove connector J1 from the A/D card. The connector plugs into a header on the A/D card (see

Figure 2-23). Wire the load cell cable from the load cell or junction box to connector J1 as shown in Table 2-2.

SIG+

SIG–

SEN+

SEN–

EXC+

EXC–

J1

JP2

JP1

Figure 2-24. Single-Channel A/D Card

If using six-wire load cell cable (with sense wires), remove jumpers JP1 and JP2 before reinstalling connector J1.

For four-wire installation, leave jumpers JP1 and JP2 on. For six-wire load cell connections on dual-channel A/D

cards, remove jumpers JP3 and JP4 for connections to J2.