2 zero the inclinometer – Rice Lake TradeRoute HL Series - Installation and Service Manual (Legal-for-Trade) User Manual

Page 30

Advertising
background image

26

Trade

Route Installation Manual

3.2

Zero the Inclinometer

SOFTKEY5

....

Exit

Zero Angle

SCALES

SERIAL

FEATURE

PORT 1

INCLIN

....

....

....

BAUD

The inclinometer has been zeroed at the factory on a level surface. Once the electronic weigh center has
been installed, perform this function with the system display level.

Figure 3-3. Inclinometer Settings

Adjust the scale platform for level in both pitch and roll using a bubble or electronic level.

1. Put the indicator in configuration mode.
2. Navigate over to

SERIAL

, then down to

PORT 1

.

3. Navigate down to

ICLIN

.

4. Navigate down to

BAUD

.

5. Press

Zero Angle

. There will be no indication the command was accomplished; check the angle indications to

verify.

6. Press

Exit

to return to normal weigh mode.

7. Once in normal weigh mode, press

Setup Menu

. The pitch and roll values will be displayed. Due to heavy

filtering it may take several seconds for the angle values to settle.