920i® Electronic Weigh Center

31

10. Use the numeric keypad to enter the actual value of the calibration test weights and press

to save the

value and go to span calibration (

WSPAN

).

11. Place test weights on scale.
12. Press

again to show the current WSPAN value.

13. Press the

Calibrate

softkey to calibrate span. When complete, the new A/D count for the span calibration is

displayed.

14. Press

again to save the span calibration value and go to the next prompt (

WLIN

).

Five-Point Linearization (Using the WLIN Parameter)

This provides increased scale accuracy by calibrating the indicator at up to five additional points between the zero

and span calibrations.

Note

Linearization is optional: if you choose not to perform linearization, skip the WLIN parameter; if linearization
values have previously been entered, these values are reset to zero during calibration. WZERO and WSPAN
must be calibrated before adding linearization points; WLIN values must be less than the WSPAN value and

must not duplicate WZERO or WSPAN.

To perform linearization, follow the procedure below:

1. With

WLIN

displayed, press

down

to go to the first linearization point (

POINT 1

).

2. Press

again to show the weight value prompt (

WGT 1

).

3. Press

once more to show the weight value.

4. Place test weights on the scale; use the numeric keypad to enter the actual test weight value.

5. Press

to save the value and move to the calibration (

CAL 1

) prompt.

6. Press

to show the current calibration value.

7. Press the

Calibrate

softkey to calibrate the linearization point. When complete, the A/D count for the linear

calibration is displayed.

8. Press

again to save the calibration value and go to the next prompt (

POINT 2

).

9. Repeat for up to five linearization points.
10. To exit the linearization parameters, press

to return to

WLIN

.

Re-zero Function

The optional re-zero function is used to remove a calibration offset when hooks or chains are used to hang the test

weights.

Note

Note

The re-zero function cannot be used with five-point linear calibration.

If no other apparatus was used to hang the test weights during calibration, remove the test weights and
press

to return to the

CALIBR

submenu.

If hooks or chains were used during calibration, remove these and the test weights from the scale. With
all weight removed, go to the

REZERO

parameter and press

to show the current zero value. Press

the

Calibrate

softkey to adjust the zero and span calibration values. Press

or

to return to the

CALIBR

submenu.

Press

to return to the SCALES menu, or press the Save and Exit softkey to exit setup mode.