Trade

Route Installation Manual

3.0

920i

®

Electronic Weigh Center

Important

Configuration and calibration should be performed only by a Rice Lake distributor for Legal for Trade
sealing.

3.1

Configuration

The 920i indicator (integral in the electronic weigh center) is set up with all required parameters when it is shipped

with

Trade

Route

system. To interact with installed hardware the common parameters requiring configuration to all

Trade

Route

systems are shown in Table 3-3. The appropriate parameter values for popular net capacities are listed

in Table 3-4 and Table 3-5. Double check that the parameters are set appropriately for your system.

SCALES

SERIAL

FEATURE

SCALE x

GRADS

10000

number

See

FORMAT

Submenu

CONFIG

FORMAT

SPLIT

3INTVL

3RNG

2RNG

2INTVL

OFF

Specify for

SPLIT = OFF

....

....

....

....

Configuration of the
displayed capacity and
resolution is required.

Note

For a more detailed summary and instructions of configuration, see the 920i Installation Manual (PN 67887).

Figure 3-1. Scales Menu

Table 3-1. Scales Menu

SCALES Menu

Parameter

Choices

Description

Level 2 submenus

SCALEx

Allows configuration and calibration of each scale.

CONFIG

Lists available and associated A/Ds.

Level 3 submenus

GRADS

10000
1–9999999

Specifies the number of full-scale graduations if SPLIT=OFF. (For multi-range and
multi-interval scales [SPLIT

OFF], the GRADS value is derived from the capacity and

display divisions specified for the range or interval.)

The value entered must be in the range 1–9999999 and should be consistent with legal
requirements and environmental limits on system resolution.

To calculate GRADS, use the formula: GRADS = Capacity / Display Divisions.

Display divisions are specified under the FORMAT submenu.

SPLIT

OFF
2RNG
3RNG
2INTVL
3INTVL

Specifies whether the scale is full-range (OFF), multi-range (2RNG, 3RNG), or multi-interval
(2INTVL, 3INTVL). See 920i Manual for further descriptions.

FORMAT

PRIMAR
SECNDR

For standard scales (SPLIT=OFF), see submenu descriptions in Table 3-2. Also see 920i
Manual for further descriptions.