Operator’s manual – ANKOM TDF Dietary User Manual

Page 15

Advertising
background image

Operator’s Manual


Rev E 12/2/14

pg. 15

10. Connect the Enzyme Containers to the corresponding ports.

The Enzyme Containers connect into labeled, color-coded ports on the left side of the instrument using push-on

connectors. Make sure that you hear a click sound when you connect the enzyme containers to the ports.

11. Connect the Nitrogen Supply to the Nitrogen Supply Port.

The Nitrogen Supply Port has a Push-to-Connect fitting that accepts
¼” OD tubing. Acceptable tubing materials include Nylon,
Polyurethane, Polyolefin, and Teflon. The tubing material must be
suitable for the specified pressure.

12. Set up the Peristaltic Pump.

Unscrew the six acorn nuts on the
back of the instrument and remove the
clear back panel to access the pump
area. You might have to tilt the
instrument

forward

into

Service

Access Position to access the top two
nuts located behind the black torsion
bar. Clip and remove the cable tie
from around the pump. Latch the two
pump doors so they are snapped
closed.

13. Reinstall the clear back panel and tighten the acorn nuts.

Push-on
Connector

Nitrogen Supply Port

100 psi (6.9 bar)


Acorn Nuts (the top two are behind the black torsion bar)

Cable Tie

Pump Doors