Line charge, Temperature monitor (°c), Operator’s manual – ANKOM TDF Dietary User Manual

Page 77

Advertising
background image

Operator’s Manual


Rev E 12/2/14

pg. 77

When you press one of the

Continue buttons, the Pump Tube Test screen is once again displayed with the bottom part of the

screen showing the status of each tube. This diagnostic is automatically executed at the beginning of each new IDF/SDF or TDF
run.

Temperatures

When you press the

Temperatures button on the Diagnostics screen, the following screen will be displayed.

The readings for the In-line Heater and all six Heating Paddles dynamically change as the temperature changes.

Line Charge

When you press the

Line Charge button on the Diagnostics screen, the following screen will be displayed.

This diagnostic allows you to pump individual fluids, or all of the fluids, into the lines from the Chemical Containers to the
Tubing Support Panel. If by accident you run out of any fluid during a normal IDF/SDF or TDF run, there may not be enough
fluid in the lines for the next assay. In that case you should use this diagnostic to re-fill the lines that ran out.

A button will change color as the associated line is being charged. A line is charged when its associated button returns to its
original green color.

Line Charge

Amylase

Protease

AMG

HCl

EtOH78

EtOH95

Water

Buffer

ALL

Temperature Monitor (°C)

XX

Paddle 1

XX

Paddle 2

XX

Paddle 3

XX

Paddle 4

XX

Paddle 5

XX

Paddle 6

XX

In-line Heater