Set valves, Motor test, Operator’s manual – ANKOM TDF Dietary User Manual

Page 75

Advertising
background image

Operator’s Manual


Rev E 12/2/14

pg. 75

Motor Test

When you press the

Motor Test button on the Diagnostics screen, the following screen will be displayed.

This diagnostic allows you to test the motor at different speeds (default is 240 RPM) and directions using different amounts of
fluid. Before pressing the

GO button to start the Motor Test, it is recommended that you first open at least one Supply Valve and

one Output Valve. To do that, press the

Set Valves button. The following screen will be displayed.

This diagnostic also displays the maximum pressure read from the vacuum sensor during the test. The value starts at zero and
will display a value while the motor is running and when it stops. This can be used for troubleshooting.

Running the Motor Test could leave fluid in some of the lines. After running a
Motor Test, always run the Clear IDF/SDF Lines diagnostic.


Set Valves

Amylase Supply

(closed)

HCl Supply

(closed)

AMG Supply

(closed)

Protease Supply

(closed)

Water Supply

(closed)

EtOH95 Supply

(closed)

EtOH78 Supply

(closed)

Buffer Supply

(closed)

IDF Inlet

(open)

Waste

(open)

SDF Inlet

(open)

Output

Supply

Motor Test

Set Valves

GO

Motor

Direction

(FWD)

120 RPM

240 RPM

Motor Speed

10.0

Amount (ml)

00.0

Max Vacuum (psi)