Check ph manually, Sdf filter minutes ok, Yes no – ANKOM TDF Dietary User Manual

Page 54

Advertising
background image

Operator’s Manual


pg. 54

Rev E 12/2/14

17. Connect the Nitrogen Supply to the instrument and turn on.

Make sure the Nitrogen supply in your lab is connected to the instrument and turned on. The pressure gauges on the
instrument should show 50-55 psi on the left and about 4 psi on the right.

Press the check mark () button on the Touch Screen Display. The following screen will be displayed.

18. Set filter times (in minutes).

Because different samples take different amounts of time to filter, the above screen allows you to set your filter times. To
change any of the times shown on the screen, press the specific gray button. A number pad will be displayed that will
allow you to enter the time that you want. The times you enter will remain until you change them again.

The initial filter times shown when you first run the instrument are based on factory
experience. During filtration the computer allows you to bypass the filter time if
you notice the filtering is complete. The computer also allows you to add time
during filtering if needed (see the “Status Screen” section of this manual for more
detail).

When all of the times shown on the screen are what you want, press the check mark () button on the Touch Screen
Display. The following screen will be displayed.

19. Set the manual pH check for after the Protease digestion.

When running the AOAC 985.29 method, the instrument will always stop just before the Protease digestion (after the
bags have cooled to the appropriate temperature) and ask you to adjust pH.

If you plan to also check/adjust the pH after the Protease digestion, press the

Yes button on the screen above. Otherwise,

press the

No button. The following screen will be displayed.

Check pH manually?

AOAC 985.29 TDF

Yes

No

AOAC 985.29 TDF

After release

After EtOH rinses

10.0

3.0

SDF Filter minutes OK?

(Press button with number to change)

Advertising