Select a function, Diagnostics – ANKOM TDF Dietary User Manual

Page 18

Advertising
background image

Operator’s Manual


pg. 18

Rev E 12/2/14

6. Weigh Diatomaceous Earth (DE).

DE is used during fiber analysis to enhance the flocculation and filtration of the SDF fraction. Place ca 1 g of DE in each
of six tared and numbered tins and record the weights.

7. Weigh Samples.

Place 0.5

0.05 g of sample in each of six tared and numbered tins and record the weights.

8. Turn the instrument power on.

When you turn the power on, the instrument will run through an initialization process and the Control Panel will turn on.

9. On the Touch Screen Display, select the instrument function you would like to perform.

The Control Panel on the ANKOM

TDF

Dietary Fiber Analyzer uses Touch Screen technology. To operate the instrument

you will press identified buttons on the Touch Screen Display and buttons below the screen.

When your instrument is initialized and ready to operate, the following screen will be displayed.

Although the instrument works in accordance with AACC and NMKL methods, the
Touch Screen Display only refers to the AOAC methods.

9.1

If this is the first time the instrument is being operated after being installed, or after sitting unused for a period of
time (see Appendix B), or after tubing was replaced, the lines must be charged to avoid a Vacuum Sensor Fault.
To charge the lines follow the steps below.

9.1.1 Press the

Diagnostics button on the “Select a Function” screen. The following screen will be displayed.

9.1.2 Press the

Line Charge button. The following screen will be displayed.

Diagnostics

Touch Screen Display version: A.AA
SPLat controller software version: vX.XX dd-mm-yy

Pump Tube

Test

Expert Mode

(OFF)

Faults

Service

Mode

Clear IDF /

SDF lines

Digestion

Times

Heater Test

Temperatures

Pressures

Line Charge

Motor Test

Valve Test

Select a Function

AOAC 991.43

IDF/SDF

AOAC 985.29

TDF

Diagnostics

AOAC 991.43

TDF

Advertising