Operator’s manual – ANKOM TDF Dietary User Manual

Page 20

Advertising
background image

Operator’s Manual


pg. 20

Rev E 12/2/14

10. Install SDF filter bags on the ANKOM

TDF

Dietary Fiber Analyzer.

10.1

Remove Clamp Bars A, B, C, and D by lifting them off of the locator rods.

10.2

Gently pull the black SDF Delivery Nozzle out toward you.

10.3

Place a labeled and weighed SDF bag up underneath the SDF Delivery Nozzle so that the Delivery Nozzle is
inside the top part of the bag. Pull the bag up so that the top of the bag is about 35 mm (1.375 inches) above the
top of Clamp Bar C and return the Delivery Nozzle to its original position. This will hold the back of the bag in
place.

As part of normal operation, solution from the IDF bag will flow into the SDF bag.
Therefore, when installing the SDF filter bags it is very important to position them
high enough vertically so that at least 20 mm (0.75 inches) of the filter part of the
IDF bag can fit inside the top of the SDF bag.

Clamp Bar C

Clamp Bar D

Clamp Bar B

Clamp Bar A