Technical data

This section introduces the reader to the instrument specifications and the list of materials used.

3 .1

Materials used

Materials used

Component

Material designation

Base plate

Aluminium / stainless steel

Wedges

TeO

2

Polarizers

Glass

Housing

Polyurethane foam

Detector

InGaAs

Magnets

NdFeB

3 .2

Technical data basic instrument

3 .2 .1

Basic instrument

Technical data basic instrument
Dimensions housing ( W x H x D)

350 x 250 x 450 mm

Electric power supply

100−230 VAC ± 10 %, 50/60 Hz, 350 W

Ambient conditions

< 80 % relative humidity for T < 31 °C, linear decreasing
to 67 % at 35 °C, max. 2500 m, for indoor use only

Ambient temperature

5−35 °C ( 25 ± 5 °C recommended)

Pollution degree

2

Installation category

II

Spectral range

800 - 2500 nm (default 1000 - 2500 nm)
12‘500 - 4‘000 cm

-1

(default 10‘000 - 4‘000 cm

-1

)

(if not specified otherwise for measuring cell)

Resolution

8 cm

-1

(with boxcar apodisation)

Type of interferometer

Polarisation interferometer with TeO

2

wedges

Wavenumber accuracy

± 0.2 cm

-1

(measured with HF gascell at an ambient

temperature of 25 °C ± 5 °C)

S/N

10000 (peak-to-peak noise of a linear corrected baseline
between 5600−6000 cm

-1

, measured with NIRFlex

Liquids, 2 x 64 scans, Blackman apodisation)

Number of scans/sec.

2−4

Analog digital converter

24 bit

Type of lamp/lifetime lamp (MTBF)

Tungsten halogen lamp / 12000 h (2 x 6000 h)

Type of laser

12 VDC HeNe, wavelength at 632.992 nm

Ethernet connection

100 Mbit/s