5 Putting into operation

54

NIRFlex N-500 Operation Manual, Version F

5 .8

Installing measurement equipment to the NIRFlex Fiber Optic SMA

To install the instrument, proceed as follows:

left connector = short spring

right connector = long spring

To mount the SMA-adapter onto the optical
fiber cable:

Put the spring onto the fiber connector

and secure the end of the connector with
the threadened spacer nut.

NOTE
The spring must be freely movable after
installation! Fasten the spacer nut handtight
only!

Insert the optical fiber cable.

Place it into the cable-slot of the stress

relief and swivel the stress relief to match
the screw heads at the SMA socket.
Push the stress relief against the spring

of the optical fiber cable and swivel it into
vertical position.
Fasten the two fixating screws of the

stress relief hand-tight.

Fixate the optical cable with cable ties.