6 .2 general recommendations for measuring liquids, 6 .3 starting a measurement – BUCHI NIRFlex N-500 User Manual

Page 59

Advertising
background image

6 Operation

59

NIRFlex N-500 Operation Manual, Version F

6 .2

General recommendations for measuring liquids

For sample measurements with the NIRFlex Liquids, make sure that the cuvettes are placed in the
direction of the centre of the cuvette holder. This is important to ensure optimal heat transfer to the
sample.
The calibration wavelength range with glass cuvettes goes from 4000 to 10000 cm

-1

.

The useable wavelength range for measurements with a fiber optic probe is 4500 – 10000 cm

-1

.

Temperature differences can cause a spectral shift. The samples should either be measured at a
constant temperature or the temperature influences must be considered during calibration.
Avoid air bubbles.
Number of scans: Qualitative calibration: 4−10; quantitative calibration: 16−32 (max 64)

General recommendation for measuring liquids

Sampling option

Substance

External Reference measurement

Comment

NIRFlex Liquids

Transparent liquids

Built-in External Reference
(air)

1, 2, 5, 10 mm
cuvette with spacer

NIRFlex Fiber Optic Liquids

Transparent liquids

aggressive, trans-

parent liquids

Clean probe head

Max. temperature
150 °C

avoid air bubbles

NIRFlex Solids with petri dish
add-on and transflectance cover

Clear liquids, turbid
liquids

Empty petri dish with
transflectance cover

Avoid air bubbles

NIRFlex Solids with flow-cell
adapter add-on

Clear liquids, turbid
liquids

Clean flow cell

Avoid air bubbles

NIRFlex Solids with Spinner add-on
and transflectance cover

Clear liquids, turbid
liquids

Empty cuvette with trans-
flectance cover

Avoid air bubbles

NIRFlex Fiber Optic Solids with
transflectance adapter

Clear liquids, turbid
liquids

Clean probe head

Make sure there
are no air bubbles
in the gap

6 .3

Starting a measurement

The spectrometer needs a warm-up time of approx. 15 min. The lamp and the laser within the instru-
ment will be switched on as soon as the NIRWare software is running.

When the instrument has warmed up and the sample is presented to the instrument, you can start the
measurement.
Generally, you have two possibilities to do so:

Press the START button at the measuring cell.

1.

Click the Start icon (green arrow) in the NIRWare Operator.

2.

NOTE
When you are working with the NIRFlex Fiber Optic Solids or NIRFlex Fiber Optic Liquids probe, you
can also use the remote button on the handle to start a measurement.