4 Description of function

42

NIRFlex N-500 Operation Manual, Version F

The LEDs at the handle of the fiber optic probe indicate the status of the system and the results of the
prediction in routine operation.

Applications without calibration (just measuring spectra, e.g. acquisition of calibration spectra)

-

green LED flashing

measuring

-

green LED on

measurement finished

red LED on

green LED on

message on monitor

Application with calibration (routine use for predictions)

-

green LED flashing

measuring

red LED on

green LED on

message on monitor

-

green LED on

result conforms

red LED on

-

result does not conform

Because the fiber optic probe is used at a certain distance from the spectrometer, the length of the
fiber optic cables between the measuring cell (spectrometer) and the probe tip is about 2 m (special
versions with longer fiber optic cables are available on request).

4

3

2

1

7

8

9

10

11

12

13

14

6

5

a Protection cover
b LEDs: OK, not OK
c START button
d Handle
e Monofiber fiber optic 2 m
f Electrical connections
g Light barrier

h Holder for fiber optic
i Internal reference
j Optic 3
k Detector
l Motor
m Optic 1
n Optic 2

Functional principle of the NIRFlex Fiber Optic Liquids