6 Operation

64

NIRFlex N-500 Operation Manual, Version F

6 .7 .2

NIRFlex Fiber Optic Liquids

For most of the applications a pathlength of 1 or 2 mm is the best choice. Many samples will contain
water or alcohol, which cause strong NIR absorptions. Therefore for the standard NIRFlex Fiber Optic
Liquids a pathlength of 2 mm will be sufficient (special versions with other pathlengths are available).
For sample measurements with the NIRFlex Fiber Optic Liquids make sure that the quartz window in
the probe head is in full contact with the sample to be analyzed.
Despite its rugged design, the transmission probe requires careful treatment.

6 .7 .3

NIRFlex Fiber Optic SMA

This measurement setup is suitable for continuous online measurements. Special settings can be
found in the NIRWare manual.

NOTE
Respect handling and operation information provided by the manufacturer of your custom optics /
accessories.