Warning, 6 .7 nirflex fiber optics, Notice – BUCHI NIRFlex N-500 User Manual

Page 63

Advertising
background image

6 Operation

63

NIRFlex N-500 Operation Manual, Version F

!

WARNING

Risk of fire by highly flammable materials

Do not insert highly flammable materials or samples into the measurment cell

6 .7

NIRFlex Fiber Optics

All fiber optics must be treated with special care to avoid breaks and other damages of the glass fibers
and optical surfaces! Cleaning should always be done manually with a soft, lint free damp cloth with
mild soapy water. Optical surfaces can be cleaned with Kimwipes.

NOTICE

Risk of damage at the fiber optics

Do not bend the fiber optics to a radius smaller than 15 cm

Do not overstrech or twist the fiber optics

Do not expose the fiber optics to heavy impacts or punctual pressure

Do not bring the fiber optics into contact with organic solvents, acids or alkaline solutions

Do not bring the white standard into contact with furanes, acids and other oxidizing agents

6 .7 .1

NIRFlex Fiber Optic Solids

The handle is equipped with a remote control for initiating the measurements remotely and displaying
the results using LEDs on the handle.
Because the fiber optic probe is used at a certain distance from the spectrometer, the length of the
fiber optic cables between the measuring cell (spectrometer) and the probe tip is about 2 m (special
versions with longer fiber optic cables are available on request).

NOTICE

Risk of damage at the white standard

Do not bring the white standard into contact with furanes, acids and other oxidizing agents

NOTE

The software automatically detects the presence of a transflectance adapter but not the path-

length in use, so make sure to use the appropriate pathlength.
It is not possible to start a reference measurement if the reference cap has not been detected by

the system.

Reference measurement

For correct measurements and correct results, it is necessary to record a reference spectrum regularly.
For this purpose, use the reference cap. In order to avoid false results, the reference cap is detected
automatically if installed.