4 Description of function

23

NIRFlex N-500 Operation Manual, Version F

4 .2 .3

Transmission mode

NIR light is sent through a defined pathlength of sample material (e.g. in a cuvette). The transmitted
light contains the spectral information. This is the preferred method for testing liquids.

Detector

Cuvette

NIR light

pathlength

4 .2 .4

Diffuse transmission mode

The diffuse transmission mode is a mixture of “diffuse reflection” and “transmission” mode.

The NIR light penetrates the sample:

is refracted

diffusely reflected

diffusely transmitted

The transmitted rays contain the spectral information of the sample.

Detector

NIR light

Sample

Collecting lens