4 Description of function

22

NIRFlex N-500 Operation Manual, Version F

NOTE

With special accessories, reflection-mode probes can be equipped for transflectance.

The samples should either be measured at a constant temperature or the temperature influences

must be considered during calibration.
Recommended number of scans:

Qualitative calibration: 4−16; quantitative calibration: 16−32

Most liquid sample materials can be constantly analyzed with the help of a flow-cell adapter.

4 .2 .2

Diffuse reflection mode

Most non-translucent materials (e.g. solids such as powders, pellets and cereals) can be analyzed
via diffuse reflection. The NIR light penetration is limited by the sample material. It interacts with the
sample, is refracted and diffusely reflected into the sensor. The reflected rays contain the spectral
information of the sample.

Detector

NIR light

Sample material

Factors that can influence the measurement:

Not enough sample material available

Sample not homogenious or representative

Humidity of sample material

Temperature of sample material

Sample cup inadeqate (e.g. cup material and thickness, blind spots, dirt etc.)

NOTE

The samples should either be measured at a constant temperature or the temperature influences

must be considered during calibration.
Coarse grained samples should be milled before carrying out the measurement. Recommended

number of scans: Qualitative calibration: 4−16; quantitative calibration: 16−32;
granules: 64.