5 .5 installing the measuring cells, Notice – BUCHI NIRFlex N-500 User Manual

Page 49

Advertising
background image

5 Putting into operation

49

NIRFlex N-500 Operation Manual, Version F

5 .5

Installing the measuring cells

All different measuring cells are provided with a plug connection.

Installing a measuring cell

To install a measuring cell, proceed as follows:

Place the measuring cell vertically into the provided space.

Carefully applying pressure to the measuring cell, make sure that the plug connection catches.

Make sure it is seated well by means of the centerning pins at the NIRFlex N-500.

Subsequently, fasten the measuring cell with the fixating screw at the bottom of the NIRFlex N-500.

NOTE
When using the NIRFlex Solids Transmittance cell, put the sample plate onto the cell by inserting the
central hole into the shaft. The guiding pin must be inserted into the corresponding hole (next to the
handle).

NOTICE

Risk of measurement interruption and data errors by communication loss.

Fixate the measuring cell to the basic instrument using the fixing screw at the bottom plate of

the instrument.