5 Putting into operation

46

NIRFlex N-500 Operation Manual, Version F

The instrument must be installed:

indoor only with a clearance of 15 cm minimum to walls and any other objects

without being exposed to heating or cooling source (e.g. direct sunlight or air conditioning)

with the transportation lock in unlocked posit

ion (see previous section)

on a stable and horizontal position

NOTE
Because of its weight it is recommended to carry the instrument with two persons.

If the instrument is installed on a laboratory cart, the cart should have large, soft wheels (approx. 10
cm diameter) for better shock absorption at transport on even floors.
To avoid damage by condensate moisture, special attention is needed when the instrument is moved
between environments with bigger temperature difference! When moving the instrument from a colder
to a warmer environment, let it acclimatize to the ambient temperature before switching it on!

NOTICE

Risk of instrument damage by moisture condensate.

Let the instrument sufficiently acclimatize at relocation

The electric plug on the measuring cell is only designed to plug-in the measuring cells. Do not connect
any other item.
The spectrometer must be placed onto a clean, level and stable support. It is recommended to set up
the spectrometer so that it is easily accessible from all sides.
The spectrometer is normally operated from the front. Make sure that ergonomic and back-friendly
operation is insured.

5 .3

Electrical connection requirements

After the pre-installation procedure (all internal connections have been established and transportation
lock is in unlocked position) has been completed successfully, the power plug of the NIRFlex N-500
can be connected to mains.

The used mains circuit has to:

provide the voltage that is given on the type plate of the instrument.

be able to handle the load of the connected instruments.

be equipped with adequate fuses and electrical safety measures, in particular proper grounding.

See also technical data of all components regarding the different minimum system requirements!

NOTICE

Risk of instrument damage by wrong mains supply.

External mains supply must meet the voltage given on the type plate

Check for proper grounding

Exchange defective cabling instantly