6 Operation

58

NIRFlex N-500 Operation Manual, Version F

6

Operation

This section describes how to start a measurement and how to prepare a reference measurement and
gives recommendations on possible applications of the instrument.

NOTICE

Risk of instrument damage at use with engaged transportation lock.

Disengage the transportation lock prior use

Keep shocks and vibrations away

6 .1

General recommendations for measuring solids

The calibration wavelength range with a standard fiber optic probe goes from 4500 to 10000 cm

-1

.

For quantitative applications the fiber optic probe should be fixed.
Hygroscopic samples show temperature dependency. The samples should either be measured at a
constant temperature or the temperature influences must be considered during calibration.
Coarse grained samples must be possibly milled before carrying out the measurement.
Number of scans: Qualitative calibration: 4−16 (max. 32); quantitative calibration: 16−32;
granules: 64. The spectral range of the NIRFlex Solids Transmittance is limited to 6000 cm

-1

. The

number of scans should be 64 by default.

General recommendations for measuring solids

Sampling option

Substance

External Reference measurement

Comment

NIRFlex Fiber Optic Solids

Homogenous powder
(qualitative) solids with
flat surface

White standard in the
reference cap

NIRFlex Solids with petri dish
add-on

Powder (quantitative)
granules

White standard in a refer-
ence holder

Make sure that there is
enough sample in the dish so
that all the light is reflected
by the sample

NIRFlex Solids with XL add-on Powder in containers

pellets

White standard on XL
add-on

Make sure that there is a
constant bulk density, a
constant sample container
quality and a reproducible
container position

NIRFlex Solids with vial
add-on

Powder in small glass
vials

White standard in vial
add-on

NIRFlex Solids with Spinner
add-on

Powder (quantitative),
granules

White standard in a refer-
ence holder

NIRFlex Solids Transmittance

tablets, capsules,
powders in small
plastic cups

Open beam (reference
position in sample plate)