7 Maintenance and repairs

71

NIRFlex N-500 Operation Manual, Version F

NOTICE

Risk of instrument damage by ingress of particles.

Check proper seating of all housing parts before use

Do not operate instrument when housing is open or unlocked

Resetting the lamp lifetime counter
After lamp module exchange, the following message appears at power on:

Message concerning required instrument setup

Click the OK icon

to confirm the message.

Now click the Service icon

(Operator) and double-click ‘

Assembly’.

NOTE
To correctly monitor the lamp lifetime the software hour counter has to be reset after each lamp
replacement. At lamp exchange, the secondary lamp module is active.

The following dialog opens:

Assembly setup dialog

Click the icon next to the Lamp row

to reset the “housing open” status.