7 Maintenance and repairs

72

NIRFlex N-500 Operation Manual, Version F

Workflow to reset the lifetime counter of the primary lamp module

To establish the standard working conditions of the NIRFlex N-500:

a

b

c

Start the Operator software

Switch from “secondary lamp”

a to “primary lamp”

Resetting the hour

b counter will bring up a prompt for confirmation

Confirm the message

c to finally set the lamp lifetime counter to zero

➡ After this procedure, reboot the NIRFlex N-500 and restart the Operator software.