NIRCal 5.5 Software Manual

122

NIRCal 5.5 Manual, Version A


If too many primary PCs are selected, a calibration is overfitted. In this case only the
calibration spectra will be identified correctly but not necessarily the validation samples in the
project.
Example of overfitting:


The residuals of the C-Set are much lower, as for the V-Set.

If too few primary PCs are selected, the spectra are underfitted, the calibration might not be
selective enough for the products.
Example of underfitting: