Introduction

NIRCal 5.5 Manual, Version A

25

It is not possible to import data from NIRCal 4 ".ncf" format, because ncf-files do not include spectral
data.
It is important to take care of the compatibility of the spectra from different instruments!

Link:

Convert and Import..in NIRWare DB

.

1.2.11 Selections

Data selections are needed to group the data before the calculation of a calibration.
For each calibration the selections can be different.

Calibration method selection

Choose the method according to the target of the application.

For quantitative applications there are MLR, PCR or PLS and for qualitative application Cluster or
SIMCA are available.

Validation method selection

To test the validity of a calibration internal test samples are used in the validation mode: VS. For
quantitative calibration the Cross Validation: CV is available.

The default method is VS: Validation Set, where the V-Set spectra should be selected by the user.
You can also use the spectra selection wizard.

Data Sets Selection

Spectra Selection

The selection can be made in NIR-Explorer:



All Spectra

not modifiable, determined through the number of imported spectra.

User Spectra

mainly used for visual comparison, any selection is acceptable, this selection
has no correlation with the C-, V- or U-Set spectra.

Calibration
Spectra

spectra selected for the calibration: C-Set, about 2/3 of all spectra for VS
mode.

Validation Spectra

spectra selected for the validation: V-Set (1/3 for VS).

Outlier Spectra

result of the "Outlier Detection" wizard for the statistical residual, score,
property value and leverage outliers in C- and / or V -Set


Each of these sets, except of "All Spectra", can be edited by the user.