Adjusting primary pcs (calculation parameters), 7 adjusting primary pcs (calculation parameters) – BUCHI NIRCal User Manual

Page 123

Advertising
background image

Chemometrics

NIRCal 5.5 Manual, Version A

123


There are spectral characteristics, which are still not fitted, the residuum spectra are bigger, as
the noise of the instrument.


NOTE
Avoid underfitting and overfitting! The PCs / Factor Selection Wizard can help with an estimate.

3.7 Adjusting Primary PCs (Calculation Parameters)

Menu:

Calibration \ Parameter \ Calculation

Icon:


The primary PCs are edited in the Dialog called Calculation Parameters. To check the result of the
Setting of the Primary Principal components open the graphic Spectra Residuum: Menu: Graphics /
Spectra /

Residuum

:


NOTE
Avoid overfitting: do not use too many principal components with little information.

Advertising