NIRCal 5.5 Software Manual

134

NIRCal 5.5 Manual, Version A

This calculations can be
made for the C-Set and/or
V-Set as selected by the
user.



The confidence interval e.g.
for 99 % level (mean value
+/- c * stdev) can be
determined according the
number of C-set spectra:
Calculate c.


Press Recalculate beside
the 4 different types of
outliers to start the
calculations. The found
outlier type spectra are
marked red after pressing
Recalculate."

By pressing Plot, the
graphic is opened with
scatter plot for easier
spectra selection.

Outliers are marked with a
sign and colored as shown
in the plot legend.


Outlier spectra are listed in the NIR-Explorer under the spectra selection of the active calibration:


Spectrum No. 72-74 are identified as residual outliers with this module (residuals: 0.0067-0.0075).