V-set spectra residuum max, V-set spectra residuum min, V-set std orig property – BUCHI NIRCal User Manual

Page 206: 94 v-set spectra residuum max, 95 v-set spectra residuum min, 96 v-set std orig property

Advertising
background image

NIRCal 5.5 Software Manual

206

NIRCal 5.5 Manual, Version A

3.18.94

V-Set Spectra Residuum Max

Description

Shows the maximum residuum of the V-Set spectra using different number of
PCs in the selected wavelength range.

Use

To find the optimum number of primary PCs.

Method

PCR / PLS / Cluster (CLU) / SIMCA

Matrices ID

77

Tip

Small residuum indicates a good reconstruction, but avoid overfitting.

Details

The residuum with e.g. PCs 3 shows the residuum spectra using 1-3 PCs for
the reconstruction.

Related Topic

Spectra Residuum

,

Primary PCs

3.18.95

V-Set Spectra Residuum Min

Description

Shows the minimum residuum of the V-Set spectra using different number of
PCs in the selected wavelength range.

Use

To find the optimum number of primary PCs.

Method

PCR / PLS / Cluster (CLU) / SIMCA

Matrices ID

78

Tip

Small residuum indicates a good reconstruction, but avoid overfitting.

Details

The residuum with e.g. PCs 3 shows the residuum spectra using 1-3 PCs for
the reconstruction.

Related Topic

Spectra Residuum

,

Primary PCs

3.18.96

V-Set Std Orig Property

Description

Standard deviation of the original property values of the validation set spectra
(standard deviation of the whole validation range, X-axis)

Use

To compare with the V-Set Std Pred Property

Method

PCR / PLS

Matrices ID

23

Tip

Gives information about the size of the original validation range.

Details

It is available in the calibration protocol (V-Set Orig.sdev).

Related Topic

V-Set Std Pred Property

Advertising