Calibration score limits, Cluster per property, Consistency – BUCHI NIRCal User Manual

Page 170: 4 calibration score limits, 5 cluster per property, 6 consistency

Advertising
background image

NIRCal 5.5 Software Manual

170

NIRCal 5.5 Manual, Version A

3.18.4 Calibration Score Limits

Description

Allowed minimum and maximum score values of the calibration.

Use

Score outlier detection in the validation set, especially in Cluster calibration.

Method

PCR / PLS / Cluster (CLU) / SIMCA

Matrices ID

45

Tip

Only scores outliers according the secondary PCs are important.

Details

Related Topic

Scores


The allowed score limits for the validation: 1.05 x min. and max. scores of the C-set spectra.
The default blow up limit of 1.05 can be edited by the user.

3.18.5 Cluster per Property

Description

Number of cluster -connected radii area- per property.

Use

Additional information for property separation.

Method

Cluster (CLU) / SIMCA

Matrices ID

52

Tip

The number of cluster per property is always 1 in a good calibration.

Details

Each spectra of the same property should be together and build one
connected cluster.

Related Topic

Property Box Radii

,

Property Box Center


A good calibration has only one cluster for each property

3.18.6 Consistency

Description

Consistency describes the relation between the standard errors of the
calibration and validation sets.

Use

Selecting the secondary PCs for a robust calibration.

Method

PCR / PLS (only with VS mode)

Matrices ID

33

Tip

SEP- and SEE-value should be small and similar for a stable calibration.
With an optimal number of PCs the consistency shows values around 100
(80-110).

Details

Visible in the Overview, in the View: Secondary PCs Selection and under
Graphics.

Formula

Related Topic

C-Set SEE

,

V-Set SEE (SEP)

Advertising