Chemometrics

NIRCal 5.5 Manual, Version A

175

3.18.20

CV SECV

Description

CV SECV: Standard Error Cross Validated

Use

The CV selects the number of secondary PCs, where CV SECV has a
minimum.

Method

PCR / PLS

Matrices ID

72

Tip

The secondary PCs selection is indicated with red color.

Details

The first local minimum of CV SECV is taken.

Formula

Related Topic

SEE

3.18.21

CV Spectra Residuals

Description

The spectra residual of the spectra in the V-Set during CV.

Use

Compare the CV spectra residual with the residual of final calibration to find
out possible outliers.

Method

PCR / PLS

Matrices ID

71

Tip

Big CV spectra residual indicates outlier.

Details

The number of primary PCs are used for the calculation.

Related Topic

Residuals

,

Primary PCs

3.18.22

CV V-Set PRESS

Description

CV V-Set PRESS is the Predicted Residual Error Sum Square for the
samples in case they are in the V-Set of the calibration.

Use

To estimate the number of PCs. Compare it with C-Set PRESS of the final
calibration.

Method

PCR / PLS

Matrices ID

69

Tip

Optimal number of PC, where CV V-Set PRESS has the minimum value.

Details

Shows the CV C-Set PRESS in function of the number of PCs.

Related Topic

C-Set PRESS

,

CV C-Set PRESS

3.18.23

C-Set BCAP-PRES

Description

PRES function of the calibration Set comparable with the DOS based BCAP
Software.

Use

To estimate the number of factors

Method

PCR / PLS

Matrices ID

21

Tip

Details

Shows the C-Set PRES (Predicted Error Sum) in function of the number of
PCs.

Related Topic

V-Set BCAP Pres