Chemometrics

NIRCal 5.5 Manual, Version A

169

1. Select Matrices in the NIR-Explorer tree
2. then select the two matrices in the NIR-Explorer
3. click the right mouse and then 2D-Scatter


NOTE
To save such a created plot a workspace can be saved and reused within other projects.

3.18.2 Calibration Residual Limits

Description

The min. and max. residual is depending on the number of primary PCs.

Use

The maximum allowed residual of the calibration is valid for the application.

Method

PCR / PLS / Cluster (CLU) / SIMCA

Matrices ID

55

Tip

The highest residual of the C-Set is used for the residual limit calculation.

Details

The allowed residual = 2 * maximum C-Set residual. 2 is the default of
residual blow up, can be edited by the user.

Related Topic

Residuals

,

Primary PCs

3.18.3 Calibration Residual Limits with Blow Up

Description

The min. and max. residual is depending on the number of primary PCs.

Use

The maximum allowed residual of the calibration is valid for the application.

Method

PCR / PLS / Cluster (CLU) / SIMCA

Matrices ID

56

Tip

The highest residual of the C-Set is used for the residual limit calculation.

Details

The allowed residual = 2 * maximum C-Set residual.
2 is the default of residual blow up, can be edited by the user. Do not use
residual blow up smaller as 1!
The minimum residual = minimum C-Set residual /2. It is not used for the
application.

Related Topic

Residuals

,

Primary PCs

,

Blow Up Limits